ivOM 9
Technical Specifications :
| Eye piece Magnification | : | 12.5x/18B |
| Objective Focus | : | f=200mm |
| Working Distance | : | 190mm |
| Magnification for Main Microscope | : | 4x, 6x, 10x,16x, 25x |
| Magnification for Assistant | : | 6x, 10x,16x, |
| Diameter of Field | : | ∮58mm, ∮38mm,∮23mm,∮14mm,∮9mm, |
| Adjustable Diopter | : | ±5D |
| Adjustable Range of Pupil Distance | : | 55mm ~ 75mm |
| Focus for Assistant Microscope | : | ≥ 30mm |
| Maximal Resolution | : | 119LP/m |
| Illumination Source | : | 12V/100W, Medical cold-reflection Halogen Lamp |
| Illumination Type | : | 6°+0 °Coaxial illumination and 20° Oblique Illumination (it can be used as slit illumination) Slit Width |
| Coaxial Illumination | : | ≥ 50000lx |
| Oblique Illumination | : | ≥ 45000 lx |
| Reaching Radius of Arm | : | 1240mm |
| Adjustable Vertical Range | : | 850mm ~ 1350mm |
| Fine Focus Speed and Range | : | ≤ 2mm/s, 50mm |
| Speed for X/Y and Range | : | ≤ 2mm/s, 50mm x 50mm |
| Input Voltage | : | AC 220V ± 10% /50HZ ±1HZ, AC 110V ±10%/60HZ±1Hz |
| Power | : | 330VA |
| Fuse | : | AC 250V T4.0A(51S-040H), AC 110V T 8.0 A (51S-080L) |
| Electrical Safety Standard | : | Executive to GB 9706.1-2007, Class I |
| Packing Volume | : | 0.786cbm, 5 cartons / 213kg |
Winning Advantages:
- Advanced Apochromatic Technology and Unique Optical Balanced Technology effectively improves the binocular Visual Fusions, strengthens in third dimension and reduces surgeon's fatigue during operations.
- Fine focusing knob is under the bottom of assistant binocular.
- Assistant's microscope can be finely focused independently of the main microscope to avoid Surgeon's work.
- 20°oblique illumination can also be used as slit illumination. Slit width is continuously adjustable for ophthalmic use.